No. 2D730-148E*O
(b) EP
This consists of one EP-ASK and a persistence memory.
EP operates in 2D-mode
and M-mode and performs echo signal processing as outlined below.
l
Magnitude
calculation
Power operations
are performed for real and imaginary input data.
l
Log compression
Log compression
is performed for the echo signal.
l
Blending
Combination
focus and other vector combination
operations
are performed.
l
Edge enhancement
Edge enhancement
is performed for the echo signal.
l
AGC
Feedback processing and echo enhancement
are performed for the echo signal.
l
Black hole and noise spike filtering
Noise appearing as black and white points is eliminated.
l
Persistence
Frame correlation processing is performed.
(cl
FP
This consists of one FP-ASIC, a buffer memory, two FPGAs, a filter coefficient
memory, and a persistence memory.
FP operates in CDI mode and MCDI mode and
performs color Doppler processing.
l
Corner turning
Real and imaginary data input from the DR is stored in the buffer memory and the
data is rearranged
when processing is performed
by the MTI filter.
l
Wall filtering
MTI filtering is performed.
FIO, which reduces motion artifacts, is also pet-formed.
l
Autocorrelation
Auto-correlation
is performed.
l
Spatial averaging
The filter to be used in spatial averaging (smoothing)
is set
4-25