Adjusting The Near/Away Slide Wire - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
3.23.2
Adjusting the near/away slide wire
(1)
Perform the procedures in subsections 3.6.1 and 3.6.4, and the steps (1) and (2) in
subsection 3.6.5 of subsection 3.6 "Disassembling
the Operating Panel".
(2)
Loosen the nut.
(3)
When locking cannot be released, loosen the adjuster to adjust the wire.
(4)
When locking cannot be performed, tighten the adjuster to adjust the wire.
(5)
After adjustment,
tighten the nut.
Nut-