User Mode; Service Mode - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(2)
Adjustment
menu
(a)
User mode
The menu switches 0
0
V@lA
on the front of the LCD monitor are used to display the user
menu and to select menu items.
The [-] switch and [+] switch are used to select the brightness, contrast, screen size, and
y curve.
User menu
Menu item
I
Default value
I
Adjustment
range
Brightness
I
30%
I
0% to 100%
Contrast
I
100%
I
0% to 100%
Size
I
1
I
1 I213
Y
2
1 I21314
* To reset the setting to the default value, press the [-] switch while holding down the [+]
switch.
(b)
Service mode
1) Opening and closing the service mode menu
Hold down both the [V] switch and the [+] switch on the front of the LCD monitor for
3 seconds to display the service mode menu (main menu).
To close the service
mode menu, select the exit key @
and press the [V] switch.
Exit key
VOA - +
Menu switches
[-] switch [+] switch
2)
Menu operation
Use the [+] and [-] switches for menu operation
Press the [+] or [-] switch to move the selection cursor to the desired function and
then press the [V] switch to set the selection.
The submenu for the selected function is displayed to show more detailed functions.
Use [+] or [-] to select the desired function and then press [VI.
The adjustment
menu with an icon is displayed.
Change the value using [+] or [-] and then press [V]
to set the new value.
7-6