No. 2D730-148E*O
(3)
Removing the monitor unit
IA
! CAUTION:
Two persons
are required
when removing
the monitor
unit.
(a)
(b)
cc>
(4
(e)
0-J
(9)
Loosen the two screws CA> and four screws cB> to remove the monitor rear cover.
Remove the support rear cover at the rear of the operating panel.
To remove the support rear cover, slide it backward with holding and pushing the
center part.
Loosen the two screws CC> to remove the pinch shield cover.
Remove the two screws CD> to remove the clamps.
Remove the signal, audio, and power cables that are connected to the monitor unit.
Remove the four M6 bolts cE> to remove the monitor unit with the boss installed.
Remove the two M4 screws cF> to remove the support front cover.