Removing the Operating Panel
3.5.1
For the SSA-770A
The operating panel must be raised using the up/down slide lever in advance.
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
Removing the handle
Refer to subsection 3.3.
Removing the monitor unit
Refer to subsection 3.4.
Removing the right side cover
(a)
Remove the two rubber caps at the bottom of the cover.
(b)
Remove the two M4 screws CA> to remove the cover.
* The cover can be removed by pulling its bottom toward you.
Removing the left side cover
(a)
Remove the two rubber caps at the bottom of the cover.
(b)
Remove the two M4 screws cB> to remove the cover.
* The cover can be removed by pulling its bottom toward you.
Removing the right side cover for the peripheral units
Remove the two M4 screws ~0
to remove the cover.
* The cover can be removed by lifting it and pulling it toward you.
Removing the left side cover for the peripheral units
Remove the two M4 screws CD> to remove the cover.
* The cover can be removed by lifting it and pulling it toward you.
Removing the transducer cable arm
Remove the two M4 bolts cE> to remove the transducer cable arm.
Removing the transducer cable hook
Remove the two M4 bolts cF> to remove the transducer cable hook.
Removing the top cover
(a)
Remove the four caps.
(b)
Remove the four M4 screws <G> to remove the cover.
Removing the support front cover
Remove the two M4 screws <H> to remove the cover.