Removing The Caster Units; For The Ssa-77Oa - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
3.20
Removing the Caster Units
3.20.1
For the SSA-770A
(1)
Removing the left side cover
(a)
Remove the two rubber caps CA> at the bottom of the cover.
(b)
Remove the two M4 screws <A> to remove the cover.
* The cover can be removed by pulling its bottom toward you.
(2)
Removing the right side cover
(a)
Remove the two rubber caps cB> at the bottom of the cover.
(b)
Remove the two M4 screws cB> to remove the cover.
* The cover can be removed by pulling its bottom toward you.
(3)
Removing the front cover (required to remove the front caster unit)
(a)
When the blank cover has been mounted, loosen the M3 screw CC> and remove the
blank cover with the printer base.
(b)
Remove the four M4 screws CD> to remove the cover.
(4)
Removing the power supply section rear cover (required to remove the rear caster unit)
Remove the four M4 screws CD> to remove the cover,
(5)
Removing the base front cover (required to remove the front caster unit)
(a)
Remove the two M4 screws cF>.
(b)
Loosen the two M4 screws cG> to remove the cover.
* The cover can be removed by pulling it toward you slightly and lifting it obliquely.
(6)
Removing the base rear cover (required to remove the rear caster unit)
Loosen the two M4 screws cH> to remove the cover.
* The cover can be removed by pulling it toward you slightly and lifting it obliquely.
3-58