No. 2D730-148E*O
3.19
Removing the PHYSIO Unit
3.19.1
For the SSA-770A
(1)
(2)
(3)
(4)
Removing the right side cover
(a)
Remove the two rubber caps at the bottom of the cover.
(b)
Remove the two M4 screws <A> to remove the cover.
* The cover can be removed by pulling its bottom toward you.
Removing the right side cover for the peripheral units
Remove the two M4 screws cB> to remove the cover.
* The cover can be removed by lifting it and pulling it toward you.
Removing the metal fitting
Remove the three M3 screws CC> to remove the metal fitting.
Removing the PHYSIO unit
(a)
Remove all the cables connected
to the PHYSIO unit.
(b)
Remove the M4 screw CD> and M4 screw cE> to remove the PHYSIO unit.
Screw CD> 1
(PTR4-IOSBNI)
Metal fitting \
Right side cover for
the peripheral units
Right side cover
-Screw
cB>
(SBPP4-12SCR)
Screw <A>
(SBPP4-12SCR)
Cap
3-56