No. 2D730-148E*O
(3)
Description of each connector
(a)
OEM AC output
Output voltage:
Total of the three connectors
250 V (when 100 VAC is input)
360 V (when 120 to 230 VAC is input)
NOTE:
When the power supply of the system is turned ON, the relay in the
transformer
tray is turned ON by the relay control signal coming from the
AC/DC tray and the AC output is also set to ON.
(b)
P41 monitor AC output
PIN No.
Signal
1
180 VAC to 220 VAC
2
1 180 VAC to 220 VAC
3
1 Protective grounding
(GND)
41
-
(c)
J51 AC output for the AC/DC tray
cln the case of the transformer
tray>
cln the case of the Trans-unit>
PIN No.
Signal
1
180 VAC to 220 VAC
2
1 180 VAC to 220 VAC
3
I
-
4
1 Protective grounding
(GND)
PIN No.
Signal
1
78 VAC to 121 VAC
2
78 VAC to 121 VAC
3
78 VAC to 121 VAC
4
78 VAC to 121 VAC
5
Protective grounding (GND)
6
Protective grounding (GND)
6-5