Physio Group Box

When this option is installed on your system, the Physio group box provides
ECG selections.
Physio Control Layout
Icon
Selection
Phys Swp
(On/Off)
(Decrease Gain)
(Increase Gain)
(Position Down)
(Position Up)
1
2
3
4
5
T1
Delay
(Change Trigger-1 Delay)
T2
Delay
(Change Trigger-2 Delay)
[ 1 ]
I N S T R U C T I O N S
F O R
Description
Specifies the physio sweep speed.
Removes the physio trace from the screen or reactivates it.
Decreases the wave amplitude of the trace.
Increases the wave amplitude of the trace.
Adjusts the trace downwards.
Adjusts the trace upwards.
Displays the ECG trace.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Enables the first trigger marker for positioning at specific points
in the cardiac cycle.
The first trigger marker is always positioned at or to the right of
the QRS complex.
Specifies the position of the first trigger marker in relation to the
QRS complex of the cardiac cycle.
The selected value determines the time, in milliseconds (ms),
from the QRS complex to the trigger marker.
Possible settings: Minimum (0) to maximum (1980) in
increments of 20 ms.
Enables the second trigger marker for positioning at specific
points in the cardiac cycle.
The second trigger marker is always positioned at least 20 ms to
the right of the first trigger's position.
Specifies the position of the second trigger marker in relation to
the QRS complex of the cardiac cycle.
The selected value determines the time, in milliseconds (ms),
from the QRS complex to the trigger marker.
Possible settings: Minimum (T1 Delay setting plus 20 ms) to
maximum (2000 ms) in increments of 20 ms.
U S E
3 S y s t e m C o n t r o l s
3 - 3 7