Environmental Requirements

EMC Note: Operating the Antares ultrasound imaging system in close proximity to
sources of strong electromagnetic fields, such as radio transmitter stations or similar
installations may lead to interference visible on the monitor screen. However, the device
has been designed and tested to withstand such interference and will not be permanently
damaged.
Relative humidity:
Temperature:
System
Array
Transducers
(except C5F1,
C7F2, EV9F4)
C5F1, C7F2, EV9F4
Maximum Altitude:
*For systems with documentation devices installed.

Protective Measures

Explosion
This product is not designed for operation in areas subject
protection:
to explosion hazards.

Maximum Physical Dimensions

Width:
610 mm
Height:
1308 mm
Depth:
915 mm
Weight:
190 kg* (410 lbs) unpacked and ready for operation
160 kg** (350 lbs) unpacked and ready for operation
*For systems with documentation devices installed.
**For systems without documentation devices installed.
[ 1 ]
I N S T R U C T I O N S
F O R
Ambient
During Operation
or Transporation
10% to 80%,
10% to 95%,
non-condensing
non-condensing
+10°C to +40°C*
-20°C to +60°C
+10°C to +40°C
-40°C to +60°C
+20°C to +40°C
-5°C to +50°C
Operation up to 3,050
5,050 meters (18,000 feet)
meters (10,000 feet)
U S E
6 T e c h n i c a l D e s c r i p t i o n
During Storage
(
)
6 - 2 7