Removing the Battery Pack

Caution: When removing the battery pack, be careful not to pull or otherwise
strain the wires in the cable connecting the battery to the ultrasound system
because doing so can damage the battery pack and the ultrasound system.
To remove the battery pack:
1. With the storage bin cover removed, locate the battery pack within the
recess on the right rear panel. The battery pack is attached to the
ultrasound system by a connector at the end of a cable. The battery is
also secured within the recess by hook and loop fastener.
2. Grasp the battery pack within the recess and gently pull the battery pack
to release the hook and loop fastener.
Removing/replacing the battery pack.
3. Support the connector housing attached to the battery cable and pull the
connector straight out of the connector receptacle to disconnect the
battery.
[ 1 ]
I N S T R U C T I O N S
F O R
U S E
2 S a f e t y a n d C a r e
1 Battery recess,
beneath the storage
bin cover
2 Battery pack
3 Hook and loop
fastener, (hook side)
4 Hook and loop
fastener, (loop side)
5 Connector housing
2 - 2 7