Initial Setup

The ultrasound imaging system is initially unpacked and installed by a
Siemens representative. Your Siemens representative will verify the
operation of the system. Any transducers, documentation devices,
accessories, and options delivered with your system are also connected and
installed for you.
Each day before you use the ultrasound system, perform the Daily Checklist
procedures described below.

Daily Checklist

Perform the following each day before using the ultrasound system:
Visually inspect all transducers. Do not use a transducer that has a
cracked, punctured, or discolored casing or a frayed cable.
WARNING: To avoid electrical shock, you must visually inspect a transducer
prior to use. Do not use a transducer that has a cracked, punctured, or
discolored casing or a frayed cable.
Discoloration Exception:
Gigasept FF, may cause discoloration of transducer housings, including the face of
the transducer. You can continue to use a transducer if it is discolored due to the use
of these specific disinfectants only.
Visually inspect all power cords. Do not turn on the power if a cord is
frayed or split, or shows signs of wear.
If your system's power cord is frayed or split, or shows signs of wear,
contact your Siemens service representative for power cord replacement.
Verify that the trackball and the DGC slider controls are clean and free
from gel or contaminants.
Once the system is powered on:
Visually check the on-screen displays and lighting.
Verify that the monitor displays the current date and time.
Verify that the transducer identification and indicated frequency are
correct for the active transducer.
[ 1 ]
I N S T R U C T I O N S
F O R
The use of the approved disinfectants, Cidex OPA and
U S E
4 S y s t e m S e t u p
[1] Instructions for Use
Approved List of
Disinfectants
Ch 2
4 - 3