Siemens ACUSON Antares Instructions For Use Manual page 59

Ultrasound imaging system
Table of Contents

Advertisement

To remove and clean the input/output panel air filter:
Caution: Do not scrub, stretch, or bend the filter, or apply heat to the filter, as
doing so could damage the filter.
1. Power off and unplug the power cord from the power outlet.
2. Open the door to the input/output panel. Locate the filter tab near the
lower door hinge.
3. Grasp the filter by its tab and pull it out of the slot.
Removing the air filter from the input/output panel.
4. Rinse the air filter with running water and allow the filter to
completely dry.
To hasten drying, you may gently shake the filter, or blot the filter with a
clean, lint-free cloth.
Caution: Do not insert a wet filter as this can damage the system.
5. Slide the air filter back into the ultrasound system, taking care to orient
the flat side of the filter towards the left and to place the edges of the
filter within the filter slot.
6. Close the door to the input/output panel.
Sliding the air filter into the air filter slot.
7. Plug the power cord into the power outlet.
[ 1 ]
I N S T R U C T I O N S
F O R
U S E
2 S a f e t y a n d C a r e
1 Input/output panel
2 Door to input/output panel
3 Input/output panel air filter
1 Air filter slot
2 Air filter, with flat
side toward left
2 - 3 7