Hitachi ChromasterUltra RS 6430 Instruction Manual page 61

Diode array detector
Table of Contents

Advertisement

(2)
Thumbscrews
Clamp screw
If the lamp energy does not recover with the cleaning
procedure mentioned (1), clean the light inlet and outlet of
the flow cell unit.
(a)
Loosen two flow cell front lid thumbscrews and
detach the flow cell front lid.
Figure 5-3 Front Cover Clamping Screw
(b)
Loosen the flow cell clamping screws and separate
the flow cell unit.
Figure 5-4 Flow Cell Unit Clamping Screw
5 - 9
5.4.1
F
l
o
w
c
e
l
l
c
Front lid