APPENDIX 2 DESCRIPTION ON CONTACT SIGNAL COMMUNICATION

1. Outline

2. Contact Signals Entering the e-line connector

The following two types of contact signals are used in the
ChromasterUltra R
S
(1)
Contact signals (START, ERROR, BUSY) entering the e-
Line connector.
These are contact signals having input/output functions
that are used in common for each unit. These signals are
respectively formed from the START (IN/OUT), ERROR
(IN/OUT), and BUSY (IN/OUT) lines.
(2)
Individual contact signals used by each unit.
These are contact signals that are input and output via 3-
pin connector. Their names and functions vary with the unit.
A name that indicates the function is given at each
connector.
 Example of output contact signal: START OUT of
Autosampler
Each unit has two e-Line connectors for connection at the rear of
the main unit, and the signals of each unit in series via the e-Line
cable, synchronism of the ChromasterUltra R
be easily achieved.
For obtaining synchronism with the L-2000, Chromaster series
and D-2500 by way of contact signals, use the cables listed
below for the connection.