Removal Of Transport Screws (Chromasterultra R 6430 Dad); Removal Of Transport Screws (Chromasterultra R 6440 Fl Detector) - Hitachi ChromasterUltra RS 6430 Instruction Manual

Diode array detector
Table of Contents

Advertisement

2.2 Removal of Transport Screws (ChromasterUltra R
Appended Fig. 1-2 Removal of Transport Screws (6430 DAD)
Remove Transport Screws from ChromasterUltra R
Array Detector.
(1)
There are two screws for transport on the bottom of the
detector. Remove them using a Phillips screwdriver and
attach the washers with the accessory short screws at that
place.
Grommet
Screws for transport
(2)
Detach a grommet on the bottom of the detector and
remove a plastic screw for transport on the bottom of the
optical base using a Phillips screwdriver. Attach the
accessory short screw and the grommet at that place.
NOTE:
Keep these screws for transport because they are
needed when the instrument is relocated.
APPENDIX 1 - 5