5.6 Replacement Parts and Consumables

Table 5-2 Replacement Parts and Consumables
Part No.
893-4755
D
lamp (see-through)
2
894-4795
Hg lamp
889-0345
DRAIN TUBE D
J821399
TIME LAG FUSE 5A
889-0810
STANDARD FLOW CELL
889-0811
HIGH SENSITIVITY FLOW
CELL (Option)
The following table lists the replacement parts and consumables
used in this instrument. The majority of these parts (such as
lamps) have a lifetime which varies significantly with usage of the
instrument. The service life of each part is based on the
assumption that the instrument operates 160 hours/month (8
hours/day), 1920 hours/year (240 days).
NOTICE: The lifetime-limited parts (replacement parts and
consumables) defined in this manual should be
renewed at the determined intervals. If the instrument
is operated without renewing a lifetime-limited part
beyond the time point of its renewal, the instrument
may become faulty to pose a problem on safety.
Part Name
NOTICE: The transmittance of quartz currently used for the
optical penetration part of a flow cell falls down to
degradation due to ultraviolet exposure. Therefore, a
flow cell has duration of life.
The grades of degradation differ depending on the
type of lamp light irradiation time, lamp light intensity,
and solvent currently used.
Degradation will become slow if lamp light irradiation
time decreases.
ChromasterUltra R
automatic light extinction function after the completion
of analysis. Please make effective use of it.
Estimated Service Life
1000 hours/6 months
100 hours/2 years
9600 hours/5 years
-
-
9600 hours/5 years
(Mobile phase: methanol)
series is equipped with the
S
5 - 13
5.6
Remarks
-
-
Length: 2000 mm,
Inner diameter: 0.8 mm,
Outer diameter: 1.57
mm, Material: PTFE
Fitting: F-120 (PEEK)
-
-
-