5.7 Lamp Replacement
(h)
Connect the lead wire connector with the connector of
the instrument.
(i)
Pull the lead wire so that the length of the folded wire
is as long as possible and fasten it with pulling the
fastening band rightward. The fastening screws
should be tightened gradually in an alternate order.
If the fastening screws cannot be tightened up with
fingers, retorque the screws using a flat-blade
screwdriver.
(j)
Insert the folded wire into the space of the right side
of the lamp house.
5 - 18