6.2 Error Messages

Error Message
A "ROM ERROR" message
was received from the
ChromasterUltra Rs 6430
. Reconnect after ending
DAD
CDS, and reactivating the unit
and PC.
A "RAM ERROR" message
was received from the
ChromasterUltra Rs 6430
. Reconnect after ending
DAD
CDS, and reactivating the unit
and PC.
A "LOG INFORMATION
ERROR" message was
received from the
ChromasterUltra Rs 6430
. Reconnect after ending
DAD
CDS, and reactivating the unit
and PC.
A "EEPROM ERROR"
message was received from
the ChromasterUltra Rs 6430
. Reconnect after ending
DAD
CDS, and reactivating the unit
and PC.
A "PARAMETER ERROR"
message was received from
the ChromasterUltra Rs 6430
. Reconnect after ending
DAD
CDS, and reactivating the unit
and PC.
A "D2 LAMP ERROR"
message was received from
the ChromasterUltra Rs 6430
. Check the Unit Status in
DAD
the control panel.
A "Hg LAMP ERROR"
message was received from
the ChromasterUltra Rs 6430
. Check the Unit Status in
DAD
the control panel.
When the ChromasterUltra Rs 6430 Diode Array Detector is
used, error messages appear on the display unit of PC. For error
messages at the time of connection, refer to the instruction
manual of CDS or maintenance software.
Description
ROM is faulty.
RAM is faulty.
Log information is wrong.
Instrument information is
wrong.
Parameter is wrong.
Failed to turn on or off D
Failed to turn on or off Hg lamp.
6 - 3
Remedy
Turn on the module power
supply again. If this error
recurs, contact the nearest
maintenance service company
authorized by Hitachi High-
Technologies.
Turn on the module power
supply again. If this error
recurs, contact the nearest
maintenance service company
authorized by Hitachi High-
Technologies.
Turn on the module power
supply again. If this error
recurs, contact the nearest
maintenance service company
authorized by Hitachi High-
Technologies.
Turn on the module power
supply again. If this error
recurs, contact the nearest
maintenance service company
authorized by Hitachi High-
Technologies.
Turn on the module power
supply again. If this error
recurs, contact the nearest
maintenance service company
authorized by Hitachi High-
Technologies.
lamp.
Turn on the module power
2
supply again. If this error
recurs, renew the D
Turn on the module power
supply again. If this error
recurs, renew the Hg lamp.