oxyCRG Display
Approximately the bottom two thirds of the Main Screen area are
covered with the three oxyCRG waveforms.
The number of realtime waves that you can display simultaneously with
oxyCRG is related to the maximum number of realtime waves that your
system is capable of displaying. Details are provided in the following
table:
Notes on
oxyCRG
3-26
Setting up your Monitor
Standard Display
4 realtime waves
6 realtime waves
8 realtime waves
The oxyCRG information is overlayed on the realtime waves at the
bottom of the Main Screen. These realtime waves disappear from
the display, but the parameters are still being monitored, and the
parameter alarms are still active.
If oxyCRG covers part of an overlap channel, the entire overlap
channel disappears from the display.
The
Freeze Waves
of the Main Screen, because the display speed is very low.
The beat-to-beat heart rate (btbHR) in the oxyCRG display will
vary from the heart rate (HR) numeric on the Main Screen, since
the HR numeric is an averaged value.
If the monitor is switched off for less than 60 seconds, oxyCRG
data is retained. However, when the power is switched back on,
you will see gaps in the oxyCRG display. The gaps correspond to
the amount of time it takes the monitor to get the first readings
after power is restored, not the amount of time that the monitor
was switched off. For this reason, you cannot properly measure
the time difference between an event which occurs before power
Main Screen
oxyCRG Display
oxyCRG + 1 realtime wave
oxyCRG + 2 realtime waves
oxyCRG + 3 realtime waves
softkey has no effect on the oxyCRG section