Philips V24CT User's Reference Manual page 242

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Viewing Patient Data
Extended Database (CMS only)
No. of Parameters
16
16
32
32
If data is stored every minute, the values displayed are the stored
averaged values of samples taken every 12 seconds for a one-minute
interval (5 samples per minute). If data is stored every 12 seconds, the
values displayed are the exact values taken at one of the 12-second
intervals.
For example, if the database has been configured for 4 hours, typical of
the OR, the data values appear at 12 second intervals. In this case, the
values are exactly as measured, not averaged.
An asterisk on the trend or graph (*) indicates (1) a manually entered
value, (2) a resampled continuously monitored value from the last
second, or (3) a value more than 10 minutes old at the time of
resampling.
Note—
Window matches the system time (time bar at rightmost position on
trend), the screen will be updated as new data is acquired. The Vital
Signs display does not update automatically when new data becomes
available. To see new data, use the right arrow key
Printed copies of the displayed data can be made (if a printer is
available) by pressing the
the
Print Graphs
Trended vital signs can be recorded (if an M1116B PLUG-IN Recorder is
available) by selecting "Vitals PLUG-IN" in the Vital Signs task window.
8-4
Trends and Calculations
Resolution
12 seconds
1 minute
12 seconds
1 minute
When the time in the upper right corner of the Graph Trends Task
Print Review
key in the Blood Review display, or the
in the Graph Trends.
Size
9 hours
48 hours
4 hours
24 hours
Print Vitals
key in the Vital Signs display,
.