Functional
The system has a module self-test which generates and displays test
Testing
waveforms and corresponding numerics for all the modules that are
Procedures
connected and turned on. The test signals are displayed for about 30
seconds and then the display returns to monitoring mode. The test signal
will be displayed only if the system self-test was successful and the
parameter is plugged into the system and turned on.
Perform the following quick system check to verify that System board,
integral rack and modules are communication properly before verifying
the individual parameter modules.
Perform a
You can verify that the System board, integral rack and modules are
Quick System
communicating properly by completing the following test.
Check
TEST - Press the ECG key on the front of the ECG module in a system
that is turned On.
RESULT - Pressing the ECG key brings up the ECG Task Window and
verifies that the System board and integral rack are communicating with
the plug-in module.
If the plug-in module Task Window, waveform or numeric is missing from
the display, the module may not be turned On. Try pushing the key on the
plug-in to see if the system is communicating with the module.
If the ECG module does not respond, perform the module self-test.
To turn a parameter On, press the following key sequence:
1. Press
Module Setup
2.
Parametr On/Off
Pts
Refernce
3.
Select Parametr
Pts
Refernce
4.
On/Off Setup
Pts
Refernce
Performance Assurance Checks
Maintenance 13-13