Operating
The following is a list of AC and battery operating instructions.
Instructions
Caution
Philips Medical Systems recommends you fully charge the batteries
before using the V24CT for the first time. Plug the monitor into AC power
(line power) for an initial charging cycle.
The batteries continuously charge when the instrument is
connected to AC power. Charging occurs whether the monitor is
turned Off or On.
AC power is indicated by the
Battery charging is indicated by the
The LED indicator flashes when the batteries first start charging,
and continues flashing until the batteries are 30 - 40% charged.
When the batteries are 30 - 40% charged, the indicator goes from
flashing to continuously on. The
on steady until the batteries are 90% charged. The
indicator will turn off when both batteries are 90% charged or when
the line power is disconnected.
The
Battery Charged
charging function has reached 90% charged, and will stay on steady
even when the batteries are fully charged. If the line power is
disconnected, this LED indicator will be turned off.
Trends/Calc
Delayed
Record
Battery
Charging
AC Power
Battery Charging
Battery Charging
green LED illuminates when the battery
Module Setup
Battery
Charged
Battery Information (V24CT and V26CT only) 12-3
AC and DC (Battery) Operation
green LED indicator.
amber LED.
indicator will be
Battery Charging
Confirm
AC
Power