Philips V24CT User's Reference Manual page 151

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Using Philips Patient Care System with an Arrhythmia Computer
Arrhythmia Alarms on the 3150A/3153A Arrhythmia Computer
Message
*** ASYSTOLE
*** V-FIB/TACH
*** V-TACH
*** TACHY yyy > xxx
*** BRADY yyy < xxx
** NON-SUSTAIN VT
** VENT RHYTHM
** RUN PVCs
** PAIR PVCs
** PACER NOT CAPT
** PACER NOT PACE
** PAUSE
** SVT
Minimum Condition Requiring an Alarm
No QRS for 4 consecutive seconds
Fibrillatory wave for 4 consecutive seconds
Consecutive PVCs >/= V-Tach Run limit and
HR > V-Tach HR limit
Heart Rate (yyy) greater than the Extreme
Tachy limit (xxx)
Heart Rate (yyy) less than the Extreme
Brady limit (xxx)
A run of Vs having a ventricular HR>V-Tach
HR limit, but lasting for less than the V-Tach
Run limit
A dominant rhythm of adjacent Vs > vent
rhythm limit and ventricular HR<V-Tach HR
limit
Run of PVCs greater than 2
Two consecutive PVCs between non-PVCs
No QRS for 1.75 x the average R-R interval
with Pace Pulse (paced patient only)
No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only)
No beat detected for 1.75 x average R-R
interval for HR <120, or no beat for 1 second
with HR >120
Run of SVPBs >/= SVT Run limit and with
SVT Heart Rate greater than the SVT HR
limit
Other Patients 4-13