Philips V24CT User's Reference Manual page 381

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Module
Cardiac Output
tcpO
/tcpCO
2
2
tcpO
2
tcpCO
2
CO
2
ETCO
2
IMCO
2
AWRR
Temperature
Performing
Auto Check - None
the ECG
Self-Test The ECG and ECG/RESP self-test may be performed at any
Module and
time as the lead selector and pace pulse detection are inhibited during the
ECG/RESP
check. The module will produce ECG test signals at the beginning of the
Self-Test
processing circuitry, and process the test signals in the same way patient
signals are processed. The test signals last about 30 seconds and then the
display returns to the normal monitoring mode.
TEST - Press
RESULT - If the module is functioning correctly, the following heart rate
and respiration are displayed without the alarms being activated:
ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode
Test Numeric
5 l/min (+0.1 l/min)
(Blood Temp: 37°C)
(Inj Temp: 0°C)
60 mmHg (8.0 kPa)
40 mmHg (5.3 kPa)
40 mmHg (6.0 kPa)
0 mmHg (0.0 kPa)
25 rpm
40°C (104°F)
Monitor Setup
Performance Assurance Checks
Test Waveform
No wave
No wave
Simulated square wave
No wave
Test Signals
.