Changing Operating Modes - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Changing Operating Modes

This function will be mainly used by service engineers for configuring the
system on installation or for servicing the system. Before you can operate
in this Task Window you have to use a password.
There are three levels of password:
Level 1 password
Level 2 password
Level 3 password allows you to select any of the four operating
Warning
Warning
The Configuration, Service and Demo modes do not support
patient monitoring. When the monitor is attached to a patient it
must be in the monitor mode.
allows you to select Monitor or Config mode only.
allows you to select Monitor or Service mode only.
modes listed below:
Monitor
This is the usual operating mode and fully supports
patient monitoring. If the system is attached to a patient
it must be in this mode.
Config
This mode is for configuring the monitor, for example,
on installation. The settings that are selected in this
mode will be the active default settings for when the
system is turned on.
Service
This mode is used by service personnel.
Demo
This mode can be used for training purposes. The data
base has 24 hrs of simulated waveform and trend data.
Changing Operating Modes
Setting up your Monitor 3-55