Philips V24CT User's Reference Manual page 187

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Channel 4
A yellow CO
alarm would produce a recording ordered as:
2
Channel 1
Channel 2
Channel 3
Channel 4
If a red P3 alarm occurred, the recording order would then become:
Channel 1
Channel 2
Channel 3
Channel 4
For alarm recordings with oxyCRG (to document neonatal events without
the Neonatal Event capability), the alarm recording items should be set to
the following:
AlRecType
All AlarmRec
In the event of a Brady, Apnea and/or Desat alarm the recorder will
document an eight minute oxyCRG recording (six minutes preceding the
event and two minutes following the event).
Respiration
ECG
Pressure 1
Pressure 2
CO
alarm
2
ECG
Pressure 1
P3 alarm
CO
alarm
2
oxyCRG
Off
Change AlarmRec to red for ECG, Resp and
SpO
.
2
Recording Functions 6-15
Alarm Recording