Perform a
The following lists how to perform the system self-test.
System Self-
Test
13-10
Maintenance
1. Examine all controls for physical condition. During the course of
this examination, ensure that the controls perform their proper
function
2. Check that the parameter module LED's light up when first
plugged in.
3. Confirm that the display lights up and displays a Main Screen.
4. Press the ECG key on the front of the ECG module in a system
that is turned On. This brings up the ECG Task Window and
verifies that the System board and integral rack are
communicating with the plug-in module. If the ECG module does
not respond, perform the ECG module Functional self-test.
5. Ensure correct operation (visually and audibly) of all alarms using,
if required, the User Guide and a patient simulator.
6. If possible connect the ECG parameter module of the unit to an
ECG simulator and verify that an adequate trace is received at
each patient lead selection.
7. Press
Monitor Setup
screen, press
Show H W Rev
Refernce
modules in the rack. Remove a module and observe the list
changes to delete the module you removed. Re-insert the module
and observe the list changes to include the module.
Monitor Revision
,
Pts
Refernce
. This provides a realtime list of the
Pts
,
. In this
Show H W Rev
Pts
Refernce