Philips V24CT User's Reference Manual page 285

Cms patient monitoring system volume 1
Table of Contents

Advertisement

19 Stored Events:
The above example of an event summary line, conveys the following
information:
Event Storage
If the number of stored events reaches a total of 44
message is displayed in the Event Review task window:
If the number of stored events exceeds a total of 48
overwritten and the following status message is displayed in the Event
Review task window:
A(D): 4(2)
Example of an Event Summary Line
A total of 19 neonatal events are stored.
Four Apnea events occurred, of which two resulted in a
Desaturation.
Three isolated Bradycardia events occurred, none of them resulted
in a Desaturation.
One associated Apnea and Bradycardia occurred which also
resulted in a Desaturation.
Nine isolated Desaturations occurred.
Two events were manually captured and stored.
"Event Storage>90% full"
"Event Storage overflow. First events overwritten"
1. 22 for V24 and V26 Patient Monitor
2. 24 for V24 and V26 Patient Monitor
Event Review
B(D): 3(0)
AB(D): 1(1)
s
s
Neonatal Event Review 9-11
Viewing Neonatal Events
D: 9
M: 2
1
, the following status
2
, the oldest events are

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

V26ctM1165M1166M1167M1175M1176 ... Show all

Table of Contents