The Test Signals Function
System Self-Test Values:
Module
ECG and ECG/
RESP
ECG
RESP
Pressure
Systolic
Diastolic
SpO
/Pleth
2
SpO
2
Pleth
tcpO
/tcpCO
2
2
tcpO
2
tcpCO
2
CO
2
ETCO
2
IMCO
2
AWRR
Temperature
3-60
Setting up your Monitor
Test Numeric Limits
100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
120 mmHg ADULT
60 mmHg PEDI/NEO
0 mmHg ADULT/PEDI/NEO
100%
Wave on display
60 mmHg (8.0 kPa)
40 mmHg (5.3 kPa)
40 mmHg (6.0 kPa)
0 mmHg (0.0 kPa)
25 rpm
40°C (104°F)
Test Waveforms
Simulated square wave and numeric
Simulated Resp wave and numeric
Simulated square wave and numerics
Numeric only
Simulated Wave
Numeric only
Simulated square wave and numerics
Numeric only