SITRANS P measuring instruments for pressure
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P Compact
for gage and absolute pressure
Function
The process pressure acts on a piezo-resistive semiconductor
measuring bridge through a remote seal and a transmission liq-
uid. The pressure transmitter converts the pressure values into a
2
load-independent current.
A compensation network makes the output signal largely inde-
pendent of the ambient temperature. As a result of a specially
adapted remote seal connection with minimized volume, the in-
fluence of the process temperature on the output signal is
greatly reduced compared to a conventional screw connection.
The pressure transmitters can be powered with a non-regulated
DC voltage of 10 to 30 V. Output signals common to measuring
technology are available.
Technical specifications
Pressure transmitters for food, pharmaceuticals and biotechnology
Mode of operation
Measuring principle
Input
Measured variable
Measured range
Output
Output signal
• Two-wire system
• Three-wire system
Measuring accuracy
Linearity error including hysteresis
(reference point adjustment)
Adjustment accuracy
Adjustment time
Influence of ambient temperature
On th enclosure
• Zero
• Measured span
On the process connection (remote
seal)
• Flange remote seal
- DN 25 / 1"
- DN 32 / 1¼"
- DN 40 / 1½"
- DN 50 / 2"
• Clamp-on seal
- DN 25 / 1"
- DN 32 / 1¼"
- DN 40 / 1½"
- DN 50 / 2"
The zero error specified for the process connection should be consid-
ered as a guideline for a standard design. We will produce a detailed
system calculation on request. Systems with reduced remote seal errors
are available on request.
2/22
Siemens FI 01 · 2009
© Siemens AG 2008
Piezo-resistive
Gage or absolute pressure
0 ... 160 mbar (0 ... 2.32 psi)
...
0 ... 40 bar (0 ... 580 psi)
4 ... 20 mA
0 ... 20 mA
To EN 60770-1
0.2% of full-scale value
0.2% of full-scale value
< 20 ms
< 0.2%/10 K of full-scale value
< 0.2%/10 K of full-scale value
Zero error (depends on design)
4.8 mbar/10 K (0.070 psi/10 K)
2.3 mbar/10 K (0.033 psi/10 K)
1.6 mbar/10 K (0.023 psi/10 K)
0.6 mbar/10 K (0.009 psi/10 K)
9.5 mbar/10 K (0.138 psi/10 K)
4.1 mbar/10 K (0.060 psi/10 K)
3.9 mbar/10 K (0.057 psi/10 K)
3.9 mbar/10 K (0.057 psi/10 K)
Rated conditions
Installation conditions
• Mounting position
Ambient conditions
• Ambient temperature
• Storage temperature
• Process temperature
• Degree of protection (to EN 60529) IP65, optional IP67
• Electromagnetic compatibility
- Emitted interference
- Interference immunity to
Design
Weight (without remote seal)
• Field housing
• Housing with plug
Housing
• Designs
• Material
Material of union nut
Process connection
• Versions
• Material of coupling
Power supply
Terminal voltage on transmitter
Rated voltage
Certificate and approvals
Classification according to pressure
equipment directive
(DRGL 97/23/EC)
Explosion protection
• Intrinsic safety "i"
- Identification
Any, vertical as standard
-10 ... +70 °C (14 ... 158 °F)
-10 ... +90 °C (14 ... 194 °F)
Max. 200 °C (392 °F), depends
on design
To EN 50081 Part 1, issue 1993
(residential and industrial areas).
The unit has no own emissions.
EN 50082 Part 2, issue March
1995 (industrial areas)
460 g ( 1.01 lb)
200 g ( 0.44 lb)
• Field housing IP65 or IP67, with
screwed gland
• Angled plug DIN 43650, IP65
• Cable connection, IP67
• Round plug connector M12,
IP65
Stainless steel, mat.
No. 1.4404/1.4305
Polyamide (with electrical con-
nection using plug or cable)
Electronics unit potted with sili-
cone
Internal ventilation for measuring
ranges < 16 bar (< 232 psi),
through housing thread or con-
nection cable depending on
design
See Ordering data
Stainless steel, mat.
No. 1.4404/316L
10 ... 30 V DC
24 V DC
For gases of fluid group 1 and liq-
uids of fluid group 1; complies
with the requirements of article 3,
paragraph 1 (appendix 1);
assigned to category III, confor-
mity evaluation module H by the
TÜV Nord
TÜV 03 ATEX 2099 X
Ex II 2G EEx ib IIC T6