SITRANS P measuring instruments for pressure
Transmitters for gage, absolute and differential pressure, flow and level
DS III series
for gage pressure
Technical specifications
SITRANS P, DS III series for gage pressure
Input
2
Measured variable
Spans (infinitely adjustable) or
nominal measuring range and
max. permissible test pressure
Lower measuring limit
• Measuring cell with silicone oil filling
• Measuring cell with inert filling liquid
Upper measuring limit
Output
Output signal
• Lower limit (infinitely adjustable)
• Upper limit (infinitely adjustable)
Load
• Without HART communication
• With HART communication
Physical bus
With polarity reversal protection
Accuracy
Reference conditions
(All error data refer always refer to the set
span)
Error in measurement and fixed-point setting
(including hysteresis and repeatability)
• Linear characteristic
- r
10
- 10 < r
30
- 30 < r
100
Long-term drift (temperature change 30 °C
( 54 °F))
Influence of ambient temperature
• at -10 ... +60 °C (14 ... 140 °F)
• at -40 ... -10 °C and +60 ... +85 °C
(-40 ... +14 °F and 140 ... 185 °F)
Measured Value Resolution
2/70
Siemens FI 01 · 2009
© Siemens AG 2008
HART
Gage pressure
Span
Max. perm. test pres-
sure

0.01 ... 1 bar g

6 bar g
(0.145 ... 14.5 psi g)
(87 psi g)
0.04 ... 4 bar g
10 bar g
(0.58 ... 58 psi g)
(145 psi g)
0.16 ... 16 bar g
32 bar g
(2.23 ... 232 psi g)
(464 psi g)
0.6 ... 63 bar g
100 bar g
(9.14 ... 914 psi g)
(1450 psi g)
1.6 ... 160 bar g
250 bar g
(23.2 ... 2320 psi g)
(3626 psi g)
4.0 ... 400 bar g
600 bar g
(58 ... 5802 psi g)
(8700 psi g)
7.0 ... 700 bar g
800 bar g
(102 ... 10153 psi g)
(11603 psi g)
30 mbar a (0.435 psi a)
30 mbar a (0.435 psi a)
100% of max. span (max. 160 bar g (2320 psi g) with oxygen measurement and inert liquid)
4 ... 20 mA
3.55 mA, factory preset to 3.84 mA
23 mA, factory preset to 20.5 mA or optionally set
to 22.0 mA
R
(U
- 10.5 V)/0.023 A in ,
B
H
U
: Power supply in V
H
R
= 230 ... 500
(SIMATIC PDM) or
B
R
= 230 ... 1100
(HART Communicator)
B
-
-
To EN 60770-1
Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm, silicone oil filling,
room temperature 25 °C (77 °F)) r: Span ratio (r = max. span/set span)
(0.0029 r + 0.071)%
(0.0045 r + 0.071)%
(0.005 r + 0.05)%
(0.25 r)% every 5 years
(0.08 r + 0.1)%
(at 700 bar:
(0.1 r + 0.2)%)
(0.1 r + 0.15)%/10 K
-
PROFIBUS PA or FOUNDATION Fieldbus
Nominal measuring
Max. perm. test pres-
range
sure
1 bar g
6 bar g
(14.5 psi g)
(87 psi g)
4 bar g
10 bar g
(58 psi g)
(145 psi g)
16 bar g
32 bar g
(232 psi g)
(464 psi g)
63 bar g
100 bar g
(914 psi g)
(1450 psi g)
160 bar g
250 bar g
(2320 psi g)
(3626 psi g)
400 bar g
600 bar g
(5802 psi g)
(8700 psi g)
700 bar g
800 bar g
(10153 psi g)
(11603 psi g)
Digital PROFIBUS PA or FOUNDATION Fieldbus
signal
-
-
-
-
IEC 61158-2
Yes
0,075%
0.25% every 5 years
0,3%
0.25%/10 K
-5
3 10
of nominal measuring range