Bar A - Siemens SITRANS P Brochure & Specs

Measuring instruments for pressure
Table of Contents

Advertisement

SITRANS P measuring instruments for pressure
Transmitters for gage, absolute and differential pressure, flow and level
DS III series for absolute pressure
(from differential pressure series)
Technical specifications
SITRANS P, DS III series for absolute pressure (from differential pressure series)
Input
2
Measured variable
Spans (infinitely adjustable) or
nominal measuring range and
max. permissible working pressure
Lower measuring limit
• Measuring cell with silicone oil filling
Upper measuring limit
Output
Output signal
• Lower limit (infinitely adjustable)
• Upper limit (infinitely adjustable)
Load
• Without HART communication
• With HART communication
Physical bus
With polarity reversal protection
Accuracy
Reference conditions
(All error data refer always refer to the set
span)
Error in measurement and fixed-point setting
(including hysteresis and repeatability)
• Linear characteristic
- r
10
- 10 < r
30
Long-term drift (temperature change
( 54 °F))
Influence of ambient temperature
• at -10 ... +60 °C (14 ... 140 °F)
• at -40 ... -10 °C and +60 ... +85 °C
(-40 ... +14 °F and 140 ... 185 °F)
Measured Value Resolution
2/98
Siemens FI 01 · 2009
© Siemens AG 2008
HART
Absolute pressure pressure
Span
Maximum working pres-
sure
8.3 ... 250 mbar a

32 bar a

(0.12 ... 3.6 psi a)
(464 psi a)
43 ... 1300 mbar a
32 bar a
(0.62 ... 18.9 psi a)
(464 psi a)
160 ... 5000 mbar a
32 bar a
(2.32 ... 72.5 psi a)
(464 psi a)
1 ... 30 bar a
160 bar a
(14.5 ... 435 psi a)
(2320 psi a)
5.3 ... 100 bar a
160 bar a
(77 ... 1450 psi a)
(2320 psi a)
(for connection thread
M10 and
the process flanges)
0 mbar a (0 psi a)
100% of max. span
4 ... 20 mA
3.55 mA, factory preset to 3.84 mA
23 mA, factory preset to 20.5 mA or optionally set
to 22.0 mA
R
(U
- 10.5 V)/0.023 A in ,
B
H
U
: Power supply in V
H
R
= 230 ... 500
(SIMATIC PDM) or
B
R
= 230 ... 1100
(HART Communicator)
B
-
-
To EN 60770-1
Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm, silicone oil filling,
room temperature 25 °C (77 °F)) r: Span ratio (r = max. span / set span)
0.1%
0.2%
30 °C
(0.1 r)%/year
(0.1 r +0.2)%
(0.1 r + 0.15)%/10 K
-
PROFIBUS PA or FOUNDATION Fieldbus
Nominal measuring
range
250 mbar a
(3.6 psi a)
1300 bar a
(18.9 psi a)
5 bar a
(72.5 psi a)
30 bar a
(435 psi a)
100 bar a
(1450 psi a)
7
/
-20 UNF in
16
Digital PROFIBUS PA or FOUNDATION Fieldbus
signal
-
-
-
-
IEC 61158-2
Yes
0.1%
0.1%/year
0.3%
0.25%/10 K
-5
3 10
of nominal measuring range
Maximum working pres-
sure
32 bar a
(464 psi a)
32 bar a
(464 psi a)
32 bar a
(464 psi a)
160 bar a
(2320 psi a)
160 bar a
(2320 psi a)
(for connection thread
7
M10 and
/
-20 UNF in
16
the process flanges)