Commission The Device; Validate The Device Measurement; Place The Device In Service - ABB XRC G5 User Manual

Hide thumbs Also See for XRC G5:
Table of Contents

Advertisement

Figure 8-23: PCCU Connection Setup for Bluetooth
Click Close to exit PCCU connection setup.
On the top PCCU menu, click Entry to connect to the XFC
When the connection is successful, continue with configuration procedures or click Close to
close the Bluetooth connection.
For further configuration or changes to the Bluetooth setup, refer to the Bluetooth help topics.
Commissioning the XFC
 Validate the device measurement
 Place the device in service.
8.8.1

Validate the device measurement

The XFC
G5
and XRC
G5
may introduce errors into the measurement results. To verify measurement accuracy, perform
pressure checks.
IMPORTANT NOTE: Save the device's data and configuration before and after calibration
procedures. For details on calibration or device data backup, see section
Validate measurement:
Perform static pressure (SP) and differential pressure (DP) calibration pressure checks.
If it is necessary to correct errors in the measurement system, proceed to section
Calibration.
8.8.2

Place the device in service

The following instructions provide general guidelines for placing the XFC
7 4 | XSE RI ES G5 | 2 10 6 02 6MNAA
or XRC
is a two-part process:
G5
G5
are calibrated at the factory before shipping. However, shipping and installation
G5
or XRC
G5
.
11
Calibration.
G5
in service.
11