Activating The Installed Software - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

GE Healthcare Software Transfer Utility that runs on a service PC. With this application,
you can transfer new software to the patient monitors over the CARESCAPE Network IX (IX
Network) or a crossover cable.
InSite RSvP
You can transfer the software in the background, without affecting normal patient monitoring.
The new, transferred software is inactive in the patient monitor(s) until you activate it.
NOTE: The software installation package contains both the patient monitor software and the
Software Transfer Utility. The software transfer procedure is described in detail in the
CARESCAPE Modular Monitors Software Installation Instructions that is delivered together with
the software installation package.

7.20.2 Activating the installed software

After transferring the new software to the patient monitors, take the transferred, inactive
software into use by activating it.
Before you start:
Verify the compatibility of the connected bedside and network devices with the new
software version. Refer to the latest version of the patient monitor's supplemental
information manual for a list of compatible network and bedside devices.
Contact GE Healthcare to get the latest version of the user and service documentation.
Contact GE Healthcare for any inquiries regarding the software installation package
and/or activation code for the new host software version.
NOTE: Loss of monitoring - Software is activated only when the patient monitor is in a patient
discharged / case reset state. Normal patient monitoring is unavailable until the software
activation is completed. This may take up to 10 minutes.
NOTE: The existing clinical and platform settings of the patient monitor are saved and are not
affected by the activation of the new patient monitor software version. However, any new or
changed clinical and platform settings in the activated patient monitor software version have
their factory default values and may require manual configuration. For more information, refer
to the latest version of the patient monitor's supplemental information manual.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window displays.
3.
In the Software List, select the software that you want to activate:
Software
Host Software
UIC Software
EMBC Software
LoFlo software
NMT software
PDM Software
PiCCO Software
sGAS Software
Purpose of use
Patient monitor
For displays and USB Remote Controller
For F5 and F7 Frames
E-LoFLo module
E-NMT module
Note: You can only upgrade to the E-NMT-01 version.
CARESCAPE Patient Data Module
E-PiCCO module
CARESCAPE respiratory modules: E-sCO, E-sCOV, E-sCOVX,
E-sCAiO, E-sCAiOV, E-sCAiOVX
Configuration
97
2106778-001

Advertisement

Table of Contents
loading

Table of Contents