Module type
Single Fault Condition (SFC)
Normal Condition (NC)
Patient (sink) leakage current tests
Electrical Safety Test completion
Functional check
Start-up
Display: Picture quality
Display: Touchscreen control
Display: Alarm light
Device Information
Configuration Information
Trim knob, keypad, and
remote:
Hard key
Trim knob, keypad, and
remote:
Trim Knob
Mouse
Alphanumeric keyboard
Barcode reader
MC Network and S/5 Network
IX printers
iPanel connection
InSite RSvP
Test completion
S/N
Ground closed (normal)
Ground open (normal)
Ground closed (reverse)
Ground open (reverse)
Polarity: Reverse
Polarity: NORMAL
Polarity: Reverse
Observed result
Observed result
Acceptance criteria
Monitor starts up normally.
Text is readable, images are clear and
brightness is good.
Touchscreen is correctly calibrated
and operates correctly.
Alarm light illuminates.
Device information is correct.
Monitor is correctly configured.
Selected menu opens and selected
activity starts.
Selected menu opens and selected
activity starts.
Selected menu opens and selected
activity starts.
Keyboard language configuration is
correct.
Data is correctly populated.
Network symbol and o
monitor's patient data is correctly
shown on the screen.
Test page is printed to the selected
printer.
The program is launched correctly.
The patient monitor is active on InSite
RSvP back office.
Installation check form
Acceptance
PASS
N.A.
criteria
 50 µA
 50 µA
 50 µA
 50 µA
10 µA
50µA
50µA
PASS
N.A.
ther
A- 3 (4)
2106778-001
FAIL
FAIL