GE CARESCAPE B850 Service Manual page 108

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
1.
In the Host Software window, verify that the software that you activate is listed as
Inactive in Current Software State.
2.
In Software License, if applicable, enter the Activation Code for the new software
version.
3.
In Software Activation, select
4.
In Erase Inactive Software After Activation, select:
No
Yes
5.
Select Activate to start the software activation.
6.
Confirm activation by selecting Yes in the Host Software window that opens.
The patient monitor shows a message Software activation after next case end / after
next discharge on the display until the next case end or patient discharge. The patient
monitoring can be continued normally until then.
7.
Software activation will start automatically after the next case end / patient discharge,
unless this process is cancelled before it starts.
The patient monitor informs clinical users about pending software activation by showing the
message Software activation after next case end / after next discharge on the display.
100
2106778-001
After next case end / discharge.
-
to keep the currently active software version as inactive software after
the new software version is activated successfully. This option lets you
restore the patient monitor to the previous software version later.
-
to erase the currently active software version permanently after the
new software version is successfully activated.