7.16.3 Testing the connectivity

NOTE: Connectivity test works only if the connection is enabled, that is the current state is
enabled/running.
1.
Log in to Webmin.
2.
Select Configuration > Remote Service.
3.
In the Sub-Modules for Remote Service menu, select Control.
4.
Below Test Connectivity in the Remote Service Control window, select Test to test the
connectivity to the remote service server.
5.
Verify that Connection to Enterprise URL and Connection to Tunnel URL both pass the
test.
6.
If one of the connection tests fail, check the failure description and fix the connectivity
issue. Re-test the connectivity.
7.17 Transferring settings from one patient monitor to
other patient monitors
You can transfer platform and/or clinical settings from one patient monitor to another, take
backup copies of the settings to an external device and restore the settings from an external
device.
1.
Complete the platform and/or clinical configuration in one patient monitor.
2.
Save the completed platform and/or clinical configuration settings to an external device
(section
3.
Load the saved platform and/or clinical configuration settings from an external device to
the destination patient monitors (section
4.
Activate the loaded platform and/or clinical configuration settings in the destination
patient monitors (section
5.
Some platform settings can't be transferred from one patient monitor to another. See
section
manually in the destination patient monitor following the instructions in sections
Configuring the
number
NOTE: Clinical settings are software package and profile specific. For example, you can transfer
clinical settings from a patient monitor with OR and PACU software only to other patient
monitors with OR and/or PACU software, not to patient monitors with ICU, NICU and ED
software.

7.17.1 Saving settings

NOTE: The patient monitor must not be configured to receive any alarm notifications from
other monitors when the settings are saved.
Before saving the settings to a patient monitor that is connected to the MC or S/5 Network,
make sure that the monitor is not receiving any alarm notifications from other monitors:
1.
Select Data&Pages > Other Patients > Receive Alarms.
2.
Check that Change All Notifications is set to Off.
Save the completed platform and/or clinical settings to a service PC or USB flash drive.
1.
Log in to Webmin.
7.17.1 Saving
settings).
7.17.3 Activating
7.17.3 Activating settings
network,
7.4 Setting unit and bed
and
7.18 License
management.
7.17.2 Loading
settings).
settings).
for details. Configure these unaffected settings
name,
7.13 Setting the host asset
Configuration
7.2