CARESCAPE Monitor B850

7.14 Password management

7.14.1 User accounts and passwords

The following table lists the available user names and default passwords for accessing the
clinical configuration, service interface, and service calibrations.
User name
biomed
clinical
For information on passwords related to other user accounts, contact GE service.

7.14.2 Changing passwords

You can change the passwords for the biomed and clinical users.
WARNING
Control of this user's password is critical to ensure that Webmin on this
device is accessed only by trained and authorized personnel. Failure to limit
access of Webmin to trained and authorized personnel only may
compromise patient safety and/or system performance.
NOTE: Username and password are case sensitive. The allowed characters for "biomed" and
"clinical" passwords are: alpha [A-Z, a-z], numeric [0-9], and space. The password length shall
be between 8 and 16 characters.
NOTE: The user name "biomed" is common for the Webmin and Service Calibrations login
screens. A change to the "biomed" password will affect both service interfaces.
1.
Log in to Webmin.
2.
Select Configuration > Passwords.
3.
In the Edit User Password window, change the biomed or clinical user's password as
required.
4.
Select Save.
The change will take effect immediately.
86
2106778-001
Default password
Access rights
Change Me
Use this user account to
access service interface
and service calibrations:
Change Me
Use this user account to
access password protected
clinical configurations:
See the user's manual and
supplemental information
manual for more
information.
Monitor Setup >
Service
Monitor Setup >
Service Calibrations
Monitor Setup >
Default Setup
Password change
Password change in
service interface is
possible for the following
user accounts:
biomed
clinical
Clinical user can change
the default password for
the clinical user account in
the clinical interface:
Monitor Setup >
Default Setup

Advertisement