CARESCAPE Monitor B850
Make sure that all the target peripheral devices are connected to the patient monitor.
NOTE: Do not disconnect any peripheral device, or shut down the patient monitor, until the
software activation is successfully completed.
To activate EMBC and UIC software:
1.
EMBC/UIC Software window, in Current Software State, verify that the software you
want to activate is listed as Inactive.
The correct UIC software depends on the display REF number:
If needed, transfer the correct UIC software before the activation.
2.
Select the peripheral devices whose software you want to activate.
3.
Select Activate to start the software activation to the selected peripheral devices.
4.
Confirm activation by selecting Yes in the Host Software window that opens.
NOTE: Software activation may take up to 10 minutes.
5.
According to your Webmin access:
If you are using Webmin on a service PC:
a.
b.
If you are using Webmin locally via the integrated browser of the patient monitor and you
are activating:
a.
b.
6.
Verify that the software activation is successful and the peripheral devices run the
activated software.

7.20.3 Canceling pending host software activation

You can cancel a pending host software activation at any time while the message Software
activation after next discharge/after next case end is shown on the display.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window opens.
3.
Select Host Software from the Software List and select
4.
In the Host Software window, below Status, verify that the Software activation pending
is Yes.
5.
Select Cancel.
102
2106778-001
M1177405 is for D15K, D19KT with REF part number 2039143-001.
2064128-001 is for D19KT with part number 2067141-001 and for D19KT VER01.
2095419-001 for D19KT with REF part number 2042925-026 and USB Remote
controller.
EMBC Software – Wait until the software activation completes and the E-Module
Frame (F5/F7) performs an automatic restart. The green On/Standby (Power) LED in
the F5/F7 frame turns off when the EMBC software activation is started and
illuminates again after the EMBC software activation is completed.
UIC Software - Wait until the software activation completes and the patient monitor
performs an automatic restart. It may take up to 10 minutes.
EMBC Software - Wait until the software activation completes and the E-Module
Frame (F5/F7) performs an automatic restart. The green On/Standby (Power) LED in
the F5/F7 frame turns off when the EMBC software activation is started and
illuminates again after the EMBC software activation is completed.
UIC Software - Select Refresh to start the activation, then select Refresh repeatedly
until the software activation progress bar shows 100%. Then wait until the software
activation completes and the patient monitor performs an automatic restart.