Installing A Psmp Module To An F5 Frame; Installing A Psmp Module To An F7 Frame; Installing Another Module To A F5 Or F7 Frame - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

6.4.2 Installing a PSMP module to an F5 Frame

1.
Connect a PSMP module by aligning it with the insertion guides on the outside of the
frame. Push the module into the frame until it stops.
2.
To remove the PSMP module, pull the pull tab out and slide the module out of the guides.

6.4.3 Installing a PSMP module to an F7 Frame

WARNING
PHYSICAL INJURY- Take care when mounting devices to an IV pole. If a
device is mounted too high the IV pole may become unbalanced and tip
over.
1.
Connect the PSMP pole mount or frame mount cable to the back of the F7 Frame.
2.
Connect the PSM module to the mount.

6.4.4 Installing another module to a F5 or F7 Frame

WARNING
Ensure that the CARESCAPE respiratory modules are in vertical position
when used. Tilting them may result in erroneous readings.
Hardware installation
61
2106778-001

Advertisement

Table of Contents
loading

Table of Contents