Connecting A F5 Or F7 Frame To A B850 Patient Monitor - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
horizontal position. To avoid the risk of equipment malfunction, make sure
the frames are not used in any other position.
WARNING
Do not use identical measurement modules or modules that map a
measurement to the same channel or parameter window. If such modules
have been connected, remove the module that has been most recently
connected. You can also remove both modules and re-connect the new
module after five seconds.
NOTE: Refer to the the patient monitor's supplemental information manual for a list of
compatible devices.

6.4.1 Connecting a F5 or F7 Frame to a B850 patient monitor

NOTE: To ensure proper grounding of the Frame, secure the cable connections by using a
screwdriver to fully seat all thumbscrews.
1.
Connect an F5 or F7 frame to ePort/Tramnet.