4.2 Local access to Webmin with a service PC

You can access Webmin locally by connecting an Ethernet crossover cable between the
service PC and the IX connector of the patient monitor.
NOTE: If you disconnect the patient monitor from a live IX Network when a patient is admitted,
you will temporarily lose the services provided by the IX Network, e.g., access to the IX printers
and MUSE reports.
WARNING
Non-medical equipment does not provide the same level of protection
against electrical shock. Do not touch the patient and any part of
non-medical equipment at the same time. Some examples of non-medical
equipment are laser printers and non-medical computers.
1.
Connect a service PC to the IX connector on the patient monitor using a crossover cable.
2.
In patient monitor, select Monitor Setup > Service. The local browser opens and displays
the Login to Webmin dialog box.
3.
Record the IX IP address of the patient monitor:
IX IP address: ______________________
IX Netmask: ______________________
NOTE: If the IX IP address field is shown as 0.0.0.0., you need to configure the patient monitor's
IX Network address first. Access Webmin using the integrated browser and configure a static IP
address for the IX Network.
4.
Configure the service PC's IP address and subnet mask to the same network segment
with the patient monitor's IX Network setting.
40
2106778-001
4.4. Login to
Webmin.
to close Webmin and return to the main display.