Installing The Battery Into The Pdm Module; Mounting The Patient Monitor; Installing The Cpu - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
One LED flashing: < 10% of full-charge capacity remaining.
NOTE: Keep the PDM connected to the patient monitor and the patient monitor connected to
the AC mains until the PDM battery is fully charged.

6.1.2 Installing the battery into the PDM module

1.
Open the battery door by gently pulling on the battery door pull tab.
2.
Pull the battery tray out of the PDM using the battery tray strap.
3.
Insert the battery with the
4.
Press the battery door closed until it seals the battery compartment.

6.2 Mounting the patient monitor

The patient monitor has an integrated GCX mounting plate. This facilitates all mounting options
for the patient monitor. Refer to the CARESCAPE Modular Monitors Mounting Solutions to
identify the compatible mounting hardware for the patient monitor.

6.2.1 Installing the CPU

WARNING
EQUIPMENT MALFUNCTION - Using the CPU in a wrong position may result
in equipment malfunction. The CPU must only be used in vertical or
horizontal position. To avoid the risk of equipment malfunction, make sure
that the CPU is not used in any other position.
WARNING
EQUIPMENT MALFUNCTION - When the CPU is used in vertical position, it is
not protected against harmful effects of dripping water (IPX1). Therefore,
54
2106778-001
TEST
button facing up and the arrow pointing into the PDM.

Advertisement

Table of Contents
loading

Table of Contents