CARESCAPE Monitor B850
Message
Identical unit & bed
name noticed
Identical
IP address
noticed
Service Monitor
Error Code
0xHOST1001
Service Monitor
Error Code
0xHOST1002
Service Monitor
Error Code
0xHOST1004
138
2106778-001
Possible causes
A patient monitor with the
identical unit and bed name is
on the network.
A patient monitor with the
identical IP address is on the
network.
One of the internal
temperature sensors indicate
the inside temperature of the
patient monitor is out of
specification. The message
stays on screen as long as the
error condition is valid.
One of the internal supply
voltages is out of the
specification. The message
stays on screen as long as the
condition is valid.
Disk usage exceeds 90%.
Possible solutions
Disconnect the patient monitor that
has the identical unit and bed name.
or
Change the unit and bed name of the
duplicate patient monitor unit and bed
name.
Disconnect the patient monitor that
has the identical IP address.
or
Change the IP address of the patient
monitor that has the duplicate IP
address.
If temperature is too high:
1. Turn off the patient monitor.
2. Let the patient monitor cool down.
3. Check that the ventilation holes are
not obstructed.
4. Ensure that the patient monitor is
installed to a location that meets
the specified environmental
requirements of operating
temperature.
5. Investigate the patient monitor
thoroughly for potential short
circuits and other electrical faults.
6. If possible, log in to Webmin and
select Diagnostics> Hardware
Stats to identify the root cause for
the error message.
If temperature is too low (or high):
If the patient monitor has been
transported or stored outside the
operating temperature range, allow it
to stabilize back to operating
temperature range before applying
power.
Log in Webmin and select Diagnostics
> Hardware Stats to identify the supply
voltage that is below or above the
specification limit.
1. Back up the clinical and platform
settings and print the licensing
information page via Webmin.
2. Re-install software.