8.3.7 Alphanumeric keyboard

Perform this test only if an alphanumeric keyboard is connected to the patient monitor.
1.
Select Data & Pages > Admit/Discharge (or Start / End Case).
2.
Select the Patient tab > Edit Name & MRN.
3.
Press
4.
Type some text into the Medical Record Number field using the connected alphanumeric
keyboard. Include some characters that are specific to the chosen keyboard locale to
verify that the keyboard language configuration is correct.
5.
Select

8.3.8 Barcode reader

Perform this test only if a barcode reader is connected to the patient monitor.
1.
Select Data & Pages > Admit/Discharge (or Start / End Case).
2.
Select the Patient tab.
3.
Scan a test barcode that is applicable to your system:
Parser Type
Length Delimited or Character
Delimited Parser
1.
2.
3.
No parser
1.
2.
3.
7. Select

8.3.9 MC Network and S/5 Network

Perform the following test only if the patient monitor is connected to a wired MC Network or S/5
Network.
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
Enter
to highlight the Medical Record Number field.
to close the open menu and return to the main display.
Test Procedure
4. Select Scan from Barcode.
5. Scan a known test barcode obtained from the hospital.
6. Verify that the data content in the barcode is correctly
4. Select Edit Name & MRN and press
5. Scan a sample barcode that only contains one piece of
6. Verify that the data is correctly populated into the
to close the open menu and return to the main display.
NOTE: The barcode data content must be known and in
compliance with the completed parser configuration.
populated to the related fields in the Patient and the
Administr. Information tabs.
the Medical Record Number field.
information (e.g., a Serial Number barcode from a
module's device label).
Medical Record Number field.
Installation checkout
Enter
to highlight