3-8 INSTALLATION

How to View or Change an HP-IB address from the Front Panel

To set an HP-IB address from the front panel, the instrument
address on the rear panel HP-IB switch (Figure 3-2) must be set to
31 (all 1s).
1. Press SYSTEM (e).
Meter Adrs , and Printer Adrs .
4. Select the desired softkey.
5. The synthesizer displays the address selected for that instrument.
6. If you want to change the address, use the keypad to enter the
desired address (0 to 30), then press @%?j.
If the synthesizer displays Rear panel HP-IB address must be 31
(11111) in order to change current address (=Xx>, the address
on the rear panel HP-IB switch (Figure 3-2) is set to something other
than 31 (all 1s).

How to Prevent a Front Panel Change to an HP-IB Address

To disable the address softkeys, set the instrument address on the
rear panel HP-IB switch (Figure 3-2) to any address
(all 1s).

How to Set the HP-IB Address on a Synthesizer without a Front Panel

If your synthesizer does not have a front panel, set the address on
the rear panel HP-IB switch (Figure 3-2) to the address you want
(factory default is 19).
All of the externally mounted connectors on the instrument are
discussed in the "Operating and Programming Reference" chapter,
under "Connectors." If you are interested in the HP part number for
a connector, see "Replaceable Parts" in this manual set.
To keep the internal timebase frequency reference oven at operating
temperature, the synthesizer must be connected to ac line power.
The synthesizer requires approximately 30 minutes to warm up
from a cold start before the OVEN display message goes off. With
a stable outside temperature, internal temperature equilibrium is
reached after approximately two hours. For additional information on
warmup times, see "Specifications," in this volume.
other than 31
Specifications