GE eBike III basic Series Service Manual page 123

Electric bike with and without nibp
Table of Contents

Advertisement

t
s
eCHnICal
peCIfICatIons e
Measuring Method:
Measuring Range
Measurement Error
Inflation Pressure
Inflation Rate
Max. Cuff Pressure
Cuff Deflation Method
Calibration
Artifact Rejection
2018111-336 Rev. B
B
III B
Ike
lood pressure Module
eBike III basic / comfort
auscultatory method, oscillometric; for resting BP, the
results from both measurements are compared for
plausibility
systolic pressure:
40 to 280 mmHg
diastolic pressure:
40 to 280 mmHg
pulse rate:
35 to 230 P/min
pressure readout error:
+/- 3 mmHg
readout resolution:
+/- 1 mmHg
300 mmHg max.; during inflation the inflation pressure
automatically adapts to patient's BP
between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
300 mmHg
pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
calibration with external pressure meter
automatic artifact rejection and comparison of the
resting BP readings from both methods for plausibility
- 123 -