The measuring equipment used for the electrical safety
tests must be inspected and calibrated at regular intervals
(refer to manufacturer of measuring equipment).
The measurement results must be documented.
B
III
/
e
Ike
BasIC
CoMfort
The eBike III basic/comfort ergometers are protection
class II equipment (no connection to protective ground).
The necessary electrical safety tests for these devices
together with the limit values are listed in the
chapter titled "eBike III basic/comfort - Preventive
maintenance"
t
s
I
eCHnICal
afety
nspeCtIons and
I
M
nspeCtIons of tHe
easurIng
The load unit is recommended to be inspected to the
approved regional standards at intervals of 2 years by a
GE Medical Systems Information Technologies authorized
service technician, and calibrated or repaired if necessary.
Calibration instructions for the NIBP module can be found
in this Service manual. Follow these instructions to display
the BP values with the test setup according to
EN 1060-1 and EN1060-3 or appropriate regional standard.
The measurement results must be documented.
2018111-336 Rev. B
s
ysteMs
eBike III basic / comfort
Warning
• Equipment Damage •
The regulations about handling ESD sensitive components
must be observed
Warning
• Patient Hazard, Equipment Damage
EN 62353 and/or IEC 60601-1 requirements must be
satisfied during repair, modifications and inspections of
medical electrical equipment
To ensure that equipment functions faultlessly and
without presenting any hazard after service interventions,
it must be subjected to the test procedures stipulated for
the different device categories
A device is considered to be unsafe, when:
-
it cannot be repaired
-
the user does not wish to have the device repaired
In this case, the operator must be informed in
writing of the hazard presented by the device
The same remark must be annotated on the
service report and on the service invoice
Warning
• Patient Hazard, Equipment Damage
Use original spare parts only
For information about product changes, refer to the
manufacturer's original documentation only
Observe the fuses' original ratings, characteristics and
Ampere interrupting capacity
Parts of the device that ensure its safe operation must be
neither damaged nor obviously unsuitable This applies to
insulation and insulating components in particular
Power cords must be visually inspected for signs of
damage before connecting them to the power line