GE eBike III basic Series Service Manual page 98

Electric bike with and without nibp
Table of Contents

Advertisement

e
CoM (e
rror Code BIts
Bit
Error
0
not used
1
not used
2
not used
3
not used
4
+24V out of range
5
not used
6
not used
7
Safety state mode
8
Setup not ready
9
CAN zykl. Timeout
- 98 -
a
xternal
nalog InterfaCe Module
Reaction
Module is switched off
Module is switched off
Module is switched off
eBike III basic / comfort
)
Display / Info
Warning
Internal error
Internal error
Internal error
Troubleshooting
- check mounting of COM
(DSUB connector)
- check jumper on LIU
- check power supply
- check connections:
power supply to LIU
- replace COM
- restart eBike (OFF / ON)
- check mounting of COM
(DSUB connector)
- check jumper on LIU
- if error persists: replace COM
- check mounting of COM
(DSUB connector)
- check jumper on LIU
- restart eBike (OFF / ON)
- if error persists: replace COM
- check mounting of COM
(DSUB connector)
- check jumper on LIU
- restart eBike (OFF / ON)
- if error persists: replace COM
2018111-336 Rev. B